The FM-9-MA2 Series: FM-9-MAB2 Air Monitors, utilize a solid-state detector and two coordinated single channel analyzer windows.
- The single channel analyzers will reject Radon pulses.
- The sample window (PHA allows the user to preset specific energy level for sample (eg. The “Radium A” peak), measuring Pu-239 or other Alpha emitters in the presence of significant or changing Radon-Thoron background.
- The count rate is shown on the front panel color monitor display both rate and integrated information.
- The user can select the operating modes: gross count, pulse height analyzing (PHA), or PHA with background subtract (PHA-SUB).
- The high-level alarm trip point is settable over the entire range.
- When activated, the system provides an on-screen red alert and an audible alarm warning.
- An adjustable alarm delay prevents short term transient alarms.
- The system uses a quick-change 47mm filter.
- The unit comes with an AP-3V air pump or may be plumbed to hood exhaust or in-house air system.
- Included with the FM-9-Air Monitor Series: Quiet Green Turtle Air Pump User’s Manual, andSix feet plastic tubing Box of 100 Filters
Specifications
For product spec information please see the attached pdf below.
Page One of Data Sheet
Page Two of Data Sheet
Complete Data Sheet
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .